• The

(unstable measurement) symbol will be shown on the display until the reading

becomes stable.

• When the reading is stable and close to the selected buffer, the (unstable measurement)

symbol will disappear and the Confirm key will become active.

• Press Confirm to confirm the calibration.
• Press ESC to exit calibration.
Point 2 calibration
• The calibrated value will be shown on the display and the second expected buffer value will

be displayed.

• Remove the electrode holder with electrodes from the top of the beaker.
• Place the second buffer into beaker and place in beaker holder. Rinse the electrodes in a

beaker containing the second buffer rinsing solution.

• Place the electrode holder (with electrodes) on the top of the beaker, lock holder by turning.
• If necessary press the ARROW keys in order to select a different buffer value.
• The

(unstable measurement) symbol will be shown on the display until the reading

becomes stable.

• When the reading is stable and close to the selected buffer, the (unstable measurement)

symbol will disappear and the Confirm key will become active.

• Press Confirm to confirm the calibration.
• The calibrated value will be shown on the display and the third expected buffer value will be

automatically selected.

• After the second calibration point has been confirmed, press ESC to exit without performing

the third calibration point.