Notes: • The calibration of the pump is dependent on concentration unit selected during SETUP.

• If an erroneous situation is encountered during the calibration, an error message is

displayed and the calibration can be restarted by pressing Restart.

• Use a new sample of standard and clean electrode and temperature probe before

pressing Restart.

• If the calibration doesn’t complete within 6 minutes the error message “Too much

standard” will be displayed and the calibration can be restarted by pressing Restart
after a new standard is prepared.