Specifications – Hanna Instruments HI 84500 User Manual
Page 5

5
Titrator
Range
Low Range: 1.0 to 40.0 ppm of SO
2
High Range: 30 to 400 ppm of SO
2
Resolution Low Range: 0.1 ppm
High Range: 1 ppm
Accuracy
Low Range: 3% of reading or ±0.5 ppm @ 25 °C, whichever is greater
High Range: 3% of reading or ±1 ppm @ 25 °C, whichever is greater
Sample volume
50 mL
Titration method Ripper method
Principle
Equivalence point redox titration
Pump speed
10 mL/min
Stirring speed
700 rpm
Log data
Up to 200 samples
ORP meter
ORP meter
-2000.0 to 2000.0 mV
ORP Resolution
0.1 mV
ORP Accuracy
± 1 mV
Log data
Up to 200 samples
ORP Electrode
HI 3148B
Environment
0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply 12 Vdc power adapter
Dimensions
235 x 200 x 150 mm (9.2 x 7.9 x 5.9”)
Weight
1.9 kg (67.0 oz.)
REQUIRED REAGENTS
Code
Description
HI 84500-50
Low Range Titrant
HI 84500-51
High Range Titrant
HI 84500-55
Calibration Standard
HI 84500-60
Acid Reagent
HI 84500-61
Alkaline Reagent (Total SO
2
)
HI 84500-62
Stabilizer Packet
SPECIFICATIONS
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)