MEASUREMENT PROCEDURE

• Refer to “Setup Menu” (see page 10) to set up the instrument for your measurement.
• For best accuracy, before taking any measurement, ensure that the pump is calibrated on the

selected range following the “Pump Calibration Procedure” (see page 19).

• Select the corresponding range according to the table below:

• Ensure the pump is primed with the correct titrant for the selected range (

HI 84500-50 Low

Range Titrant or

HI 84500-51 High Range Titrant).

Sample preparation: Use a clean pipette to precisely add the appropriate amount of wine
sample to a clean 100 mL beaker as indicated below:

e

g

n

a

R

w

o

L

)

e

l

p

m

a

s

L

m

0

5

(

e

g

n

a

R

h

g

i

H

)

e

l

p

m

a

s

L

m

0

5

(

m

p

p

0

.

0

4

o

t

0

.

1

m

p

p

0

0

4

o

t

0

3

Low Range (Free & Total SO

2

) - 50 mL

High Range (Free & Total SO

2

) - 50 mL

Note: •The volume of wine added is critical to the measurement

accuracy. Pipettes are recommended.

•Failure to use a clean pipette will result in erroneous

readings.

• Fill the 20 mL beaker up to the 5 mL mark with the

HI 84500-60

Acid Reagent and add the contents to the 100 mL beaker.

• Add the contents of one packet of

HI 84500-62 Stabilizer

Packet to the 100 mL beaker.

• Press

Titrator.

• This screen appears when the stirrer is not

working properly. Check the stir bar and beaker
content. Press

Restart to try again.

• This error message appears when the pump is

not working properly. Check the tubing, valve
and syringe. Press

Restart to try again.