Hanna Instruments HI 84500 User Manual

Page 23

background image

23

• At the end of the titration the instrument displays the concentration in ppm of SO

2

. The titration

curve can be viewed by pressing

Plot ON. Press Plot OFF to exit this mode.

• Press

LOG to record the concentration value and the titration curve into the instrument’s

memory. A message will be displayed for a few seconds indicating the amount of free log
space. Up to 200 log samples can be recorded in the instrument’s memory.

• Press

Restart to begin a new titration or ESC to return to the titration menu.

• If the concentration exceeds the range limits (>40.0 ppm for

Low Range, >400 ppm for

High Range), the exceeded range limit will be displayed blinking. Another titration can be
started by pressing

Restart.

“Wrong input” error message appears when the input reading (mV) exceeds the specified
limits. The mV value and the concentration will blink indicating an error.

• This screen appears when the stirrer is not working properly. Check the stir bar and beaker

content. Press

Restart to try again.