• The instrument will continuously update the concentration on the display. The value will be

displayed blinking. When the reading is under range “

----” symbol appears blinking.

• The titration curve can be visualized during a titration by pressing

Plot ON. Press Plot OFF to

exit this mode.

Note: DO NOT PLACE THE TIP INTO THE SAMPLE BEAKER.

PLACE THE TIP OVER A WASTE BEAKER. A SMALL
AMOUNT OF TITRANT IS DISPENSED WHEN THE PUMP
RESETS.

• Press

Start to begin a titration.

• Wait for the syringe to refill.
• Place the stir bar in the beaker and put the beaker into the beaker

holder.

• Place the probe holder on the top of the beaker and secure it by

turning clockwise.

• Rinse the ORP electrode with deionized water and immerse into

the sample until the PTFE reference junction is completely
submerged. Be sure that the tip of the electrode is not hitting the
stir bar.

• Insert the dosing tip into the titrant tube sleeve. IT IS CRITICAL

THAT THE TIP BE IMMERSED APPROXIMATELY 0.25 CM (0.1”)
INTO THE SOLUTION BEING TITRATED.

• Press

Continue to begin the titration and Stop to abort it.