• After the second calibration point is confirmed, immerse the ISE electrode

approximately 2 cm (¾”) into the third calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the Confirm

function key is displayed.

• Press Confirm to confirm calibration.
• The calibrated value and the fourth expected standard value will be

displayed.

• After the third calibration point is confirmed, immerse the ISE electrode

approximately 2 cm (¾”) into the fourth calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the Confirm

function key is displayed.

• Press Confirm to confirm calibration.
• The calibrated value and the fifth expected standard value will be dis-

played.

• After the fourth calibration point is confirmed, immerse the ISE electrode

approximately 2 cm (¾”) into the fifth calibration solution.

• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the Confirm

function key is displayed.

• Press Confirm to confirm calibration. The instrument stores the calibration

value and returns to normal measurement mode.

Note: The instrument will automatically skip the standard solutions used

during calibration.