FOUR, THREE or TWO-POINT CALIBRATION
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL or ESC after the appropriate accepted calibration point. The

instruments will return to measurement mode and will memorize the
calibration data.

SINGLE-POINT CALIBRATION
There are two selectable options for a single-point calibration:

Replace and Offset.
This option is configured in SETUP mode under the parameter First Point

Mode. In both cases the meter will alter the present calibration data in the

instrument. If “Replace” is selected a new calibration point will be added to

the existing data, and the slope is calculated. The slopes between current

buffer and nearest lower and higher buffers will be reevaluated.
If the “Offset” option is selected, an electrode offset correction is performed to

all buffer data keeping the existing slopes unchanged.
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL or ESC after the first calibration point was confirmed. The

instruments will memorize the single-point calibration data and will

return to measurement mode.

Notes: • Press MTC key to toggle between pH buffer selection and the

temperature reading during calibration while temperature probe

is disconnected (MTC mode).

• The displayed arrow is moving to the temperature value. Use

ARROW keys in order to change the temperature.