Hanna Instruments HI 3220 User Manual

Page 17

background image

17

• When the reading is stable and close to the selected buffer, the Confirm

function key is displayed.

• Press Confirm to confirm calibration.
• The calibrated value and the third expected buffer value will be dis-

played.

• After the second calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 3 cm (1¾”) into a third buffer
solution and stir gently. The temperature probe should be close to the pH
electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the Confirm

function key is displayed.

• Press Confirm to confirm calibration.
Repeat this procedure with two additional pH buffers to encompass the entire
sample pH range.