Hanna Instruments HI 3220 User Manual

Page 16

background image

16

• Press CAL. The instrument will display the measured pH, the LCD first

expected buffer and the temperature reading.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, Confirm

function key is displayed.

• Press Confirm to confirm first point.
• The calibrated value and the second expected buffer value is then dis-

played on the LCD.

• After the first calibration point is confirmed, immerse the pH electrode and

the temperature probe approximately 3 cm (1¾”) into the second buffer
solution and stir gently. The temperature probe should be close to the pH
electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.