50

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

The temperature user calibration menu can be accessed at the meter startup by pressing simultaniously three
keys as in the below drawing:

Note: The temperature user calibration is performed in three points: around 0°C, 50°C, 100°C.

To start temperature user calibration:
• Insert the probe in the beaker with water at 0°C.
• Press

to start the temperature calibration. Adjust the

temperature preset value using

or

when

necessary.

• Wait to stabilize and then press

to confirm the

calibration point.

• Repeat the above steps for 50°C and 100°C.
• Save the calibration.
• Press

to return to measure mode.

Note: Press

if you want to clear the temperature user calibration.