Temperature Compensation
The user can choose from the following options:

Linear -

the meter will compensate automatically the conductivity using the following formula:

where:

C

ref

- conductivity at reference temperature

C

i

- measured conductivity (uncompensated)

- compensation coefficient

T - temperature
ref - reference temperature

Non-Linear -

recommended for measuring the conductivity of the natural water in accordance with the

compensation table on the next page.

Disabled -

the meter will display the conductivity with no temperature compensation.

To set the

temperature compensation

mode:

• Use

or

to highlight the

Temperature

Compensation

option.

• Press

and then use

or

to select

Linear, Non-Linear

or

Disabled

option.

• Press

to confirm your selection or press

to cancel

operation.