Compensation Coefficient

(only for linear temperature compensation)

To set the compensation coefficient:
• Use

or

to highlight the

Compensation

Coefficient

option.

• Press

and set the desired compensation coefficient

using

or

to increase / decrease the value.

• Press

to save the current value or press

to

cancel operation.

Calibration / Cell Constant
The conductivity probe can be calibrated using the conductivity standards or by entering the cell constant of the
probe by the user.

Using standard solutions

The meter can be calibrated in a single or multi-points (up to four points), using 6 Hanna standards (84

µS

, 1413

µS

, 5.0

mS

,12.88

mS

, 80.0

mS

, 111.8

mS)

or using the custom standards.

The following options are available for calibration:

Standard Recognition

The user can choose between

Automatic

recognition (from 6 Hanna

standards available) or

User Standard

(when custom standards are

used for calibration).
• Use

or

to highlight the

Standard Recogni-

tion

option.

• Use

or

to choose from

Automatic

or

User

Standard

option

.

• Press

to confirm your selection or press

to

cancel operation.