SALINITY CALIBRATION

SALINITY CALIBRATION

SALINITY CALIBRATION

SALINITY CALIBRATION

SALINITY CALIBRATION

Salinity calibration is a one-point procedure at 100.0% NaCl. Use the HI 7037L calibration solution
(sea water solution) as a 100% NaCl standard solution.
To enter salinity calibration:
• Set the meter for salinity range;
• Select the

Percent Scale

(see

Salinity Setup

section);

• Rinse the probe with some of the calibration solution or deionized water;
• Immerse the probe into HI 7037L solution. The sleeve holes must be completely submerged. Tap

the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

• Enter in calibration mode by pressing

;

• Wait to stabilize;
• Press

to finish salinity calibration or press

to cancel calibration.