beautypg.com

Hanna Instruments HI 98180 User Manual

Page 56

background image

56

• The ion charge (2 chars)
• The ISE unit (2 chars)
• The short name of the selected language (3 chars)

NSLx

Requests the number of logged samples (4 chars).

The command parameter (1 char):

• P - request for pH range
• M - request for mV and Rel mV ranges
• I - request for ISE range

LODPxxx Requests the xxx

th

pH record logged data.

LODMxxx Requests the xxx

th

mV/Rel mV record logged data.

LODIxxx

Requests the xxx

th

ISE record logged data (

HI 98184, HI 98185).

LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.
LODIALL Requests all ISE Log on demand (HI 98184, HI 98185).

The answer string for each record contains:

• The logged mode (2 chars):

• 00 - pH range (0.001 resolution)
• 01 - pH range (0.01 resolution)
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range
• 05 - ISE range

• Reading status (1 char): R, O, U
• Calculated reading, with sign and decimal point and expo-

nent (11 chars) - for pH, Rel mV and ISE range

• Temperature reading, with sign and two decimal points

(7 chars)

• mV reading status (1 char): R, O, U
• The mV reading, with sign and decimal point (7 chars)
• The logged time,

yymmddhhmmss (12 chars)

• The calibration slope, with sign and decimal point (7 chars)

- not available for Rel mV range

• The calibration offset, with sign and decimal point (7 chars) -

not available for ISE

• Temperature probe presence (1 char)

This manual is related to the following products: