• After the third calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer, the

CFM

functional key is displayed.

• Press

CFM to confirm calibration.

• The calibrated value and the fifth expected buffer will be displayed.

• After the fourth calibration point is confirmed, immerse the pH electrode and the

temperature probe approximately 4 cm (1½”) into the fifth buffer solution and
stir gently. The temperature probe should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer, the

CFM

functional key is displayed.