• Press

CFM to confirm calibration.

• The calibrated value and the fourth expected standard value will be

displayed.

• After the third calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the fourth calibration solution.

• If necessary, press the

ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the

CFM

functional key is displayed.

• Press

CFM to confirm calibration.

• The calibrated value and the fifth expected standard value will be dis-

played.

• After the fourth calibration point is confirmed, immerse the ISE electrode

approximately 4 cm (1½”) into the fifth calibration solution.

• If necessary, press the

ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the

CFM

functional key is displayed.

• Press

CFM to confirm calibration. The instrument stores the calibration value

and returns to normal measurement mode.

Note: The instrument will automatically skip the standard solutions used

during calibration

FOUR, THREE, TWO or ONE-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press

ESC or CAL key after the appropriate accepted calibration point. The

instruments will return to measurement mode and will memorize the calibration
data.

ERROR SCREENS

Wrong standard
The calibration cannot be confirmed.

The message appears if mV input is out of ±2000 mV range.