WORKING WITH CUSTOM BUFFERS
If at least one custom buffer was set in SETUP menu, it can be selected for
calibration by pressing the

ARROW keys. The Custom functional key will be

displayed.

Press

Custom if you want to adjust the buffer value according with current

temperature.
Use the

ARROW keys to change the buffer value.

Press

Accept to accept new value or ESC to exit changing mode.

Note: Custom buffer value can be adjusted in a ±1.00 pH window, around

the set value.

WORKING WITH MILI pH BUFFERS
If calibration is invoked from mili pH range, the calibration buffer can be
modified in a ±0.020 pH range in according with the label on the calibra-
tion buffer.

Press

Change to enter buffer adjust mode.

Use

ARROW keys to change the buffer value.

Press

Accept to accept new value or ESC to exit adjusting mode.