Hanna Instruments HI 98180 User Manual

Page 27

background image

27

It is recommended to calibrate the instrument frequently, especially if high
accuracy is required.

The ISE range should be recalibrated:
• Whenever the ISE probe or ion charge is changed.
• At least once a week.
• After testing aggressive chemicals.
• When calibration alarm time out is expired-

“CAL DUE” tags blinks (if

feature is enabled).

Due to electrode conditioning time, the electrode must be kept immersed a few
seconds to stabilize. The user will be guided step by step during calibration
with easy to follow tags on the LCD. This will make the calibration a simple and
error-free procedure.

PROCEDURE
Select the proper ISE probe in SETUP menu (only for

HI 98185) or select the

proper Ion Charge (see SETUP for details, page 33).

Note: If ISE probe is not calibrated in at least one point, the “----” will be

displayed.

Pour small quantities of the buffer solutions into clean beakers. If possible, use
plastic beakers to minimize any EMC interferences.

For accurate calibration and to minimize cross-contamination, use two beakers for
each standard solution. One for rinsing the electrode and one for calibration.

The instrument offers a choice of six memorized standard solutions: 0.1, 1, 10,
100, 1000, 10000 ppm and calibration up to five points. For fluoride
electrode the 2 ppm standard is also available (

HI 98185 only).

Remove the protective cap from the ISE electrode.

FIVE-POINT CALIBRATION

• Immerse the ISE electrode approximately 4 cm (1½”)

into the less concentrated standard solution and stir
gently.

IIIIIS E

S E

S E

S E

S E CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

(HI 981

(HI 981

(HI 981

(HI 981

(HI 98184 &

84 &

84 &

84 &

84 & HI 981

HI 981

HI 981

HI 981

HI 9818 5

8 5

8 5

8 5

8 5)))))

This manual is related to the following products: