Page 15

background image

15

Enter the ISE mode by pressing

RANGE until the display changes to ISE

range.

Submerge the ISE electrode tip (4cm/1½”) into the sample
to be tested and wait a few seconds for the reading to
stabilize.

The ISE reading will be displayed along with the current
temperature reading.

The

“ATC” (or “MTC”) message is turned off because ppm readings are not

temperature compensated.

In order to take accurate ISE measurements, make sure that the appropriate ISE
electrode type and ISE unit were set in SETUP menu (

HI 98185 only), or the

proper ion charge is set (

HI 98184 only), and the instrument was calibrated

(see ISE CALIBRATION for details, page 27).
Notes: • When the reading is out of range, the display will flash the closest

full-scale value.

• The instrument will display

“----” on the primary LCD if it is not

calibrated. Perform at least a one-point calibration in order to
take ISE measurements.

• Changing the ISE electrode or the ion charge will need ISE range

calibration.

TEMPERATURE MEASUREMENTS

Connect the

HI 7662 temperature probe (HI 98180/1/2) or the temperature

connector (

HI 98183/4/5) to the appropriate socket. Immerse the temperature

probe (

HI 98180/1/2) or pH electrode (HI 98183/4/5) into the sample and

allow the reading on the secondary LCD to stabilize.

Note: The temperature can be displayed in Celsius degrees (ºC) or in Fahrenheit

degrees (ºF) (see SETUP for details, page 33) (not for

HI 98180).

BACKLIGHT FEATURE

The instrument is provided with a Backlight feature, which can be easily
toggled on and off through the keyboard by pressing

LIGHT.

Note: The backlight automatically shuts off after a set period (see SETUP

for details, page 33) with no buttons pressed.