• Press

CFM to confirm calibration.

• The instrument stores the calibration values and returns to normal mea-

surement mode.

FOUR, THREE or TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.

• Press

CAL or ESC after the appropriate accepted calibration point. The instru-

ments will return to measurement mode and will memorize the calibration data.

ONE-POINT CALIBRATION

Two

SETUP selectable options are available for one-point calibration:

Replace and Offset.
If the

Replace option is selected, the slopes between current buffer and

nearest lower and higher buffers will be reevaluated.

If the

“Offset” option is selected, an electrode offset correction is performed

keeping unchanged the existing slopes.
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press

CAL or ESC after the first calibration point was confirmed. The

instruments will memorize the one-point calibration data and will
return to measurement mode.

Notes: • Press MTC or MODE key to toggle between pH buffer selection and

the temperature reading during calibration while temperature
probe is not connected (MTCmode)

• The displayed arrow is moving to the temperature value. Use

ARROW keys in order to change the temperature.