Testing Fibers in Advanced Mode

MAX-700

71

Setting Pass/Fail Thresholds

The following table provides the default, minimum and maximum
thresholds.

Once the thresholds are set, the application will be able to perform
Pass/Fail tests to determine the status of the various events (pass or fail).

The Pass/Fail test is performed on two occasions:

when analyzing or reanalyzing a trace

when you open a trace file

Values that are greater than the predefined thresholds are displayed in
white on a red background in the events table.

The Pass/Fail threshold LED, located on the front of the unit, will also
indicate the status (green for pass, red for fail).

You can also set the application to display pass/fail messages when the
Pass/Fail test is performed.

Test

Default

Minimum

Maximum

Splice loss (dB)

1.000

0.015

5.000

Connector loss (dB)

1.000

0.015

5.000

Fiber section attenuation (dB/km)

0.400

0.000

5.000