beautypg.com

EXFO MAX-700 User Manual

Page 327

background image

Maintenance

MAX-700

315

Verifying Your MAX-700

To determine the measurement range (singlemode models only):
1.
Connect the OTDR as indicated below. Other configurations are

possible, but the fiber should have several sections longer than 2 km,
with no loss greater than 8 dB and with the average attenuation not
exceeding 1 dB/km. A variable attenuator will be used to adjust the loss
in the span.

One or several non-reflective events with a nominal loss of 0.5 dB
should be present. Join a series of fiber reels between the OTDR and
the variable attenuator for a length of approximately 20 km. Join
another series of reels to complete the fiber length needed for the test.

Ensure that the OTDR port and connectors are correctly cleaned.

Make sure the fiber settings are accurate (IOR, Helix factor, and
RBS).

1 to M reels of fiber

( Add as needed )

OTDR

2 km or more

Fiber End

Measured
Fiber Event

4 km min

2 km or more

20 km

Variable attenuator

2 km or more

2 km or more

Fiber end

1 to M reels of fiber

(Add as needed)

4 km min.

Measured fiber end