beautypg.com

EXFO MAX-700 User Manual

Page 334

background image

Troubleshooting

322

MAX-700

Solving Common Problems

The application displays

a message indicating

that a “live fiber error”

occurred and the fiber
was not connected to

the SM Live port.

Light has been

detected on the OTDR

port during the

acquisition or while

you were monitoring a

fiber in real-time

mode.

Disconnect the fiber from the OTDR

port. Press OK to close the message.
Start another acquisition without any

fiber connected to the OTDR. The

message about live fiber error should

not appear and the OTDR trace

should look “normal”.
If you still see the message about live

fiber error even if no fiber is

connected to the OTDR, contact

EXFO.
Never connect a live fiber to the

OTDR port without a proper setup.

Any incoming optical power ranging

from –65 dBm to –40 dBm will affect

the OTDR acquisition. The way the

acquisition will be affected depends

on the selected pulse width.

Any incoming signal greater than

– 20 dBm could damage your

MAX-700 permanently. For live-fiber

testing, refer to the SM Live port

specifications for the characteristics

of the built-in filter.

Problem

Cause

Solution