ENMET MX-32 User Manual
Page 14

ENMET Corporation
MX32
E
NGUARD
10
4.2 Display at start-up
When power is supplied to the
MX32 C
ONTROL
,
the unit goes through a self-test sequence as follows:
1.
Current software version and buzzer test.
Examples of Display:
120
ver
1 0
m32
2.
Current menu access code.
Example of Display:
1270
See section 5.0 for Maintenance Menus
Section 5.5 for Access Code
3.
Current AL1 threshold displayed for channel 1, while the AL1 and fault light for channel 1 light up.
Example of Display:
▲
▲
20
co
1
4.
Current AL2 threshold displayed for channel 1, while the AL2 and fault light for channel 1 light up.
Example of Display:
▲
▲
40
co
2
5.
Current AL1 threshold displayed for channel 2, while the AL1 and fault light for channel 2 light up.
Example of Display:
▲
▼
19.5
o2
1
6.
Current AL2 threshold displayed for channel 2, while the AL2 and fault light for channel 2 light up.
Example of Display:
▲
▲
23.5
o2
2
7.
Count down (in seconds), time for the sensor(s) to stabilize, and both fault LED are lit.
Example of Display:
0.40
min
This time is programmed at the factory
for best results. See section 5.2 to change
programmed setting.
8.
Then the
MX32 C
ONTROL
will begin the normal display cycle
Examples of Display:
▲
0
co
▲
20.9
o2
See section 4.4 for
Normal Display Cycle