0 replacement part numbers, 0 specifications – ENMET ISA-44-2OD User Manual
Page 26

ENMET Corporation
ISA-44-2-OD
22
6.0 Replacement Part Numbers
6.0 Replacement Part Numbers
ENMET replacement part numbers:
Description
Part Number
Explosionproof Sensor
03033-109
Explosionproof Sensor
03033-812
Explosionproof Sensor
03033-813
Replacement Oxygen Cell
67013-008
Replacement lens, Red ½
″
62012-011
Replacement lens, Green ½
″
62012-012
Replacement lens, Amber ½
″
62012-013
Replacement Lamp ½
″
63001-002
Replacement lens, Red 1
″
62012-001
Replacement lens, Green 1
″
62012-002
Replacement lens, Amber 1
″
62012-003
Replacement, LED Red 1
″
52006-003
Replacement, LED Green 1
″
52006-001
Replacement, LED Amber 1
″
52006-002
Replacement Oxygen Cell, PCB
03258-012
Cylinder of Calibration Gas
Contact ENMET for part number of
target gas for each instrument.
See note below.
N
OTE
: The gases should be identical, or equivalent (correlation gas) to those used to initially calibrate the unit, unless
you want to recalibrate to a new gas or different concentration. In addition, calibration gases must be in a
background of air; do not use gases with an inert gas background (such as nitrogen or argon).
7.0 Specifications
7.0 Specifications
These are general characteristics; actual specifications will vary according to custom-built features, boxes, etc.
Power
•
115 +/- 10% V
AC
single phase
•
220 V
AC
single phase, 50-60 Hz, 40 watts
•
12 V
DC
, 40 watts
Relay Current
5 amps, non-inductive surge, 2 amps steady
Sensor Life
Up to 3 years in fresh air
Oxygen
Average 14 months, 6 month warranty
Electronics
Low power, all solid-state with integrated circuits.
Audio Alarm
2900 Hz, 95 dB at 21 feet
Enclosure
NEMA-4X box, watertight, oiltight, corrosion proof (optional NEMA-12 steel or
stainless steel box).
Size
18" tall x 14.5" wide x 8" deep
Weight
48 lbs. (including cables, sensor housing, horn, etc.).
N
OTE
: All specifications stated in this manual may change without notice.