ISA-44-2-OD

ENMET Corporation

15

7. Put the cal cup of the calibration fixture over the sensor. Let the gas flow over the sensor until the alarm triggers.

Use Table 6 for the approximate allowable time period until alarm:

Table 7: Approximate Alarm Time

Gas or Vapor

Time to Alarm

20 ppm CO

5 - 7 minutes

50 ppm CO

5 minutes

200 ppm CO

2 minutes

10% LEL methane

1 minute

20% LEL methane

1 minute

100 ppm vinyl chloride

1 - 2 minutes

typical hydrocarbon vapors

1 - 2 minutes

20% LEL propane

1 minute

8. After you verify the alarm response, turn the gas off and remove the calibration fixture from the gas cylinder.

9. Turn the switch back to

OPERATE

.

C

AUTION

: If steps 1 - 10 do not trigger the alarms, and no electrical or mechanical malfunction is apparent, you must

recalibrate the instrument.