ENMET ISA-44-2OD User Manual
Page 21
ISA-44-2-OD
ENMET Corporation
17
N
OTE
: If amber light is on prior to
POT
adjustment, adjust this
POT
CW
until light is off, then readjust
CCW
until light just
activates.
I
MPORTANT
: The meter may not display the exact low level alarm.
D
O
N
OT
change the low level alarm meter reading unless grossly off (four meter needle widths or more). If far off,
recalibrate the high alarm, meter reading (step 9).
15. Repeat step 11.
16. Repeat step 12.
17. Reconnect the sample head to the air line.
18. The flow of clean air should clear the unit from its gas alarm state within 3-5 minutes. After 5 minutes, turn the
operation switch to O
PERATE
.
19. Clean compressed air lines will show a meter reading from 0 – 1/3 full scale. Green power light will indicate a clean
condition.
20. Carbon Monoxide Detection Channel Calibration is complete.
Verify the operation switch is in the O
PERATE
position.
- Formaldemeter htV (14 pages)
- PPM Formaldemeter™ htV-m (19 pages)
- PGD2 (34 pages)
- PGD2Manual.pdf (28 pages)
- RECON/4 (10 pages)
- RECON-4 (17 pages)
- RECON-IS (15 pages)
- RECON/B SERIES (17 pages)
- RECON Series (16 pages)
- OMNI-4000 (72 pages)
- QUADRANT (26 pages)
- SMARTLOGGER (19 pages)
- SPECTRUM (32 pages)
- SPECTRUM CO-RAL (18 pages)
- SPECTRUM ON-LINE (30 pages)
- SPECTRUM-RAL (15 pages)
- SPECTRUM-RAL-DC (17 pages)
- SPECTRUM SP (20 pages)
- TARGET (36 pages)
- TDX Series (8 pages)
- TX-2000 (24 pages)
- AM-5150 (23 pages)
- AM-5175 (21 pages)
- ENG – 97D STAND-ALONE (12 pages)
- GSM-60 (39 pages)
- ISA-60M with MRI-5175 (28 pages)
- MRI-5175 (2 pages)
- MEDAIR 2200 (40 pages)
- PROAIR 2200 (40 pages)
- CP-60 (23 pages)
- EX-5100 (18 pages)
- EX-5175-EC (16 pages)
- ISA-200-RAL (O) (24 pages)
- ISA-40 (19 pages)
- ISA-40M (18 pages)
- ISA-44-RALE-OD (38 pages)
- ISA-44-RAL-OD (28 pages)
- ISA-M (15 pages)
- ISA-RAL-M (22 pages)
- MedAir 2000 (30 pages)
- CD-1300-ST (13 pages)
- EX-5120 (18 pages)
- EX-5130 (16 pages)
- EX-5150-MOS (19 pages)