beautypg.com

2 calibration – ENMET ISA-44-2OD User Manual

Page 20

background image

ENMET Corporation

ISA-44-2-OD

16

5.2 Calibration

5.2.1 Toxic Gas Channel # 1

The following procedure is for checking the calibration level and for recalibrating the Carbon Monoxide(CO) Channel of
the ISA-44RAL-OD. Refer to Figure 2 and 4.

Perform A Complete Recalibration Periodically (at least monthly) and:

When you replace the sensor.

When the sensor encounters severe conditions (i.e., gross exposure to smoke or liquid hydrocarbon).

M

ATERIALS

Cylinders of calibration gas of known CO concentration in air (i.e. 20 ppm and 50 ppm CO in air.).

D

O

N

OT

use CO in an inert carrier gas, such as nitrogen or argon.

Calibration gas fixture (with quick-disconnect fitting – ENMET part number 03605-001)

Small screwdriver (for potentiometer adjustment)

dc voltmeter

P

ROCEDURE

1. Turn operation switch to H

ORN

O

FF

. Be sure humidifier bottle is filled with tap water to Water Level mark.

2. Turn operation switch to P

URGE

O

N

. Measure purge voltage across TB-6(+) and TB2-7(-). It should be 1.6 V

DC

+/-

.03 V

DC

. If not 1.6 V

DC

+/- .03 V

DC

, adjust Purge Adjust Potentiometer (see Figure 4) until desired reading is

shown. Turn operation switch to H

ORN

O

FF

and measure sensor heater voltage across TB2-6 and TB2-7. It should

now read 0.86 V

DC

+/- .03 V

DC

. If it is not 0.86 V

DC

+/- .03 V

DC

, adjust Heater Adjust Potentiometer (see Figure 4)

until desired reading is shown.

3. Turn operation switch back to P

URGE

O

N

. Purge at least one hour, up to overnight.

R

EMEMBER

: If you suspect contaminated air in your compressor system, use clean bottled air to supply the sensors with a

continuous airflow.

4. After purging, turn operation switch to O

PERATE

. Allow 30 minutes to stabilize.

5. Remove the compressed air line and insert the calibration apparatus (see Figure x).
6. Turn operation switch to H

ORN

O

FF

. Open the regulator on the top of the high-level calibration gas cylinder. This

brings the gas into the sampling assembly.

7. Regulate the flow of calibration gas according to the "...

SET

" mark or between 7 – 12

PSI

on the pressure regulator

dial.

8. Let the high-level calibration gas flow into the system for 5 minutes. This assures a full and balanced response from

the sensor. During this time, the meter should have risen above the low-level alarm point and the amber light should
have gone on.

9. With the gas still flowing, adjust the Meter Adjust Potentiometer with a small screwdriver (adjust this pot

CCW

to

increase,

CW

to decrease meter reading). You want to have the same meter reading as the calibration gas

concentration. This sets the alarm point. Refer to Figure 4.

Note:

Do not adjust for instantaneous alarm when test gas is first released. If you do, the calibration will be
inaccurate, and the sensor will be overly sensitive.

10. Adjust the High Level Alarm Set Adjust Potentiometer

CCW

until the red alarm light is just barely triggered. At this

point, verify the audio alarm is operational by turning the operation switch to O

PERATE

momentarily.

N

OTE

:

If red alarm light is on prior to the potentiometer adjustment, adjust this

POT

CW

until the light deactivates,

then readjust

CCW

until the light just barely activates.

11. Shut off the calibration gas flow (at the regulator on the gas cylinder).

12. Allow pressure to equalize – wait until the pressure meter on the humidifier-regulator assembly reads zero. Remove

calibration fixture, and reconnect humidifier-regulator assembly to compressed air line. Allow air to flow over
sensor for 10 minutes to clear the sensor.

13. In a similar manner, apply the low level calibration gas for five minutes

14. While the gas is still flowing, adjust the Low Alarm Set Adjust potentiometer

CCW

until the amber alarm light just

activates.