ENMET PROAIR 2200 User Manual
Page 30
![background image](/manuals/736131/30/background.png)
P
RO
A
IR
2200
ENMET Corporation
27
5.4.4
B High Cal/SpanGas Adjust
A High Cal/Span Gas should only be preformed after a successful Low Cal/ZeroCal has been completed.
Press the
SELECT
switch, that moves the cursor one digit to the right when the last digit is accepted the display will move
to "
HiCal xx
" gas calibration.
xx
= the level of gas to be used for calibration. The mV reading is shown in the upper right
hand corner of the display.
Apply calibration gas to sensor. See
Figure 7. After about 1 minute and mV reading has stabilized.
Press the
SELECT
switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is
successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration
Menu
Repeat above steps for each channel to be calibrated.
An unsuccessful calibration will result in an error message being displayed,
Example:
Offset Err, Cal Slope Err
If any
Err messages appear the channel needs to be re-calibrated.
To continue on too next section Press the
OPTION
switch.
Press
OPTION
switch until “Exit maint menu” appears and then press
SELECT
switch to return the instrument to the
Operational Display
When calibration is completed exit the calibration menu and maintenance menu. At the Operational Display if no error
messages are present then calibration is complete.
5.5 Dew Point Sensor
Unlike the CO and O
2
sensors, the dew point sensor cannot be field calibrated. To assure correct performance, the dew point
sensor should be replaced every two years.
5.6 Flow Control Orifice / Pre-filter
A 0.005-inch diameter orifice is used to set the flow rate and to drop the air pressure. A pre-filter is in line to help prevent
clogging. In well-maintained medical air systems, this orifice should not clog. However, if difficulty is experienced in
maintaining flow rate with assured inlet pressure, remove air pressure from the equipment and examine this orifice and pre-
filter; replace it if necessary.
Orifice location will depend on instrument sensor configuration. Orifice and pre-filter are located at the output point of the dew
point probe manifold or on the inlet line to the gas sensor manifold.