After entering a valid password to maintenance menu, see
Section 5.2.1, the calibration section is the first menu section; enter
by pressing the
SELECT
switch.
Supply sensor with clean air for LowCal/ZeroCal setting and apply calibration gas for HiCal/SpanGas setting.
Press the
SELECT
switch "
Calibration Select XX
" is displayed.
XX
= the gas to be calibrated
Press the
OPTION
switch, if needed, to change to the gas to be calibrated.
Press the
SELECT
switch, the gas & current reading are displayed in upper portion of display. The mV reading & "
LowCal
0
" is displayed in the lower portion of display. This is the LowCal setting,
usually zero, clean air must be supplied to the
sensor. This reading needs to be at or near zero. If it is not then a cylinder of clean 20.9 air should be used. See Figure 7 if
this is required.
Press the
SELECT
switch, that moves the cursor one digit to the right when the last digit is accepted the display will move
to "
HiCal xx
" gas calibration.
xx
= the level of gas to be used for calibration. The mV reading is shown in the upper right
hand corner of the display.
Apply calibration gas to sensor. See
Figure 7. After about 1 minute and mV reading has stabilized.
Press the
SELECT
switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is
successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration
Menu
An unsuccessful calibration will result in an error message being displayed,
Example:
Offset Err, Cal Slope Err.
If any
Err messages appear the channel needs to be re-calibrated.
Repeat above steps for each channel to be calibrated.
To continue on too next section Press the
OPTION
switch.
Press
OPTION
switch until “Exit maint menu” appears and then press
SELECT
switch to return the instrument to the
Operational Display
When calibration is completed exit the calibration menu and maintenance menu. At the Operational Display if no error
messages are present then calibration is complete.
Example: Full Calibration Flow Chart, for CO
From Valid Password Entry
N
OTE
:
The dew point sensor/probe can not be calibrated in the field and should be replaced every two years.
O
S
S
Calibration
Select (Gas)
MAINTENANCE MENU
Calibration
O
Press
O
PTION
until the gas to be
Calibrated is displayed
CO: XX
11
LowCal: 0000
S each digit
CO: XX
14
HiCal: 0000
S each digit
Default Calibration Points
Gas
LowCal HiCal
CO
0
20
O
2
N/A
20.9
CO
2
0
1000
HC
0
10
O = Press Option