Table 4: factory set gas alarms delay – ENMET PROAIR 2200 User Manual
Page 23
P
RO
A
IR
2200
ENMET Corporation
20
5.3.4 Set Alarm Delay
The alarms may be set to delay by 1 second increments, up to 255 seconds. Alarm delays are factory set to 5 seconds.
To change an alarm delay, you must enter the maintenance menu. Press the
OPTION
switch until “Enter Maint Menu” is
displayed then press
SELECT
switch for the Enter Password menu. Enter the valid password as described in
Section 5.2.1.
See
Table 4 below for factory set delays. A space is provided to record changes.
After entering a valid password:
Press the
OPTION
switch until;
“Maintenance Menu Set Alarm Delay”
appears on display.
Press the
SELECT
switch, "
ALARM Delay Select: XX
" is displayed.
XX
= the gas alarm to be changed.
Press the
OPTION
switch until, desired gas is displayed.
Press the
SELECT
switch; "
ALARM Delay = 0000
" is displayed, with the underscore cursor under the left digit.
Press the
OPTION
switch to change the left digit; select the correct digit.
Press the
SELECT
switch to lock in the correct digit and move the cursor one digit to the right. When the last digit is
accepted display returns to the "
Set Alarm Delay
" position.
Repeat for each sensor alarm delay to be changed.
Press
OPTION
switch until “Exit maint menu” appears and then press
SELECT
switch to return the instrument to the
Operational Display
Example: Set Alarm Delay Flow Chart
Table 4: Factory Set Gas alarms Delay
Gas
Delay
CO
5 sec
DP
5 sec
O
2
5 sec
CO
2
5 sec
O
S
MAINTENANCE MENU
Set Alarm Delays
S
O
S
Alarm Delay -
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
Press Option until the channel to be Set is displayed
Select: XX
Alarm Delay -
0005
O = Press Option
S = Press Select