beautypg.com

ENMET PROAIR 2200 User Manual

Page 24

background image

P

RO

A

IR

2200

ENMET Corporation

21


5.3.5 Relay Configuration

To change a relay configuration you must enter the maintenance menu. Press the

OPTION

switch until “Enter Maint Menu”

is displayed then press

SELECT

switch for the Enter Password menu. Enter the valid password as described below.

In the "

Enter Maint Menu

" position

Press the

SELECT

switch "

Enter Password

0

" is displayed. Press

SELECT

switch once, to move cursor to next digit,

this will be the first digit of the password.

In the

000

position, the underline cursor is under the left digit.

Press the

OPTION

switch to change the left digit; select the correct digit.

Press the

SELECT

switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the
display is transferred to the "

Calibration

" portion of the menu. If an invalid password is inserted you are returned to the

Enter Maint Menu display.

After entering a valid password:

Press the

OPTION

switch until “Configure Alarms” is displayed

Press the

SELECT

switch to enter the Configure Alarms menu

Press the

OPTION

switch to set relay configuration as needed, see below for indications

L

= Low Alarm,

H

= High Alarm,

B

= Both Alarms,

❚ = No Relay linked to channel

Press the

SELECT

switch to lock setting and move to next, channel and relay

Press

OPTION

switch until “Exit maint menu” appears and then press

SELECT

switch to return the instrument to the

Operational Display

Example: Set Relay Configuration Flow Chart

The table below shows the default relay links.

Channel 1

Channel 2

Channel 3

Channel 4

Relay 1

Low Alarm

Relay 2

Low Alarm

Relay 3

Low Alarm

Relay 4

Low Alarm

Relay 5

High Alarm

High Alarm

High Alarm

High Alarm

Relays can be linked to specific alarms.

N

OTE

:

Each operating channel must be linked to at least 1 relay.

O

S

MAINTENANCE MENU

Configure Alarms

Ch 1,2,3,4
R1 L

Pressing

O

PTION

changes letter indicated by underscore cursor

Pressing

S

ELECT

Locks underscored digit and moves cursor to next digit

Ch 1 2 3 4
R2 L

Ch 1 2 3 4
R3 L

Ch 1 2 3 4
R4 L

Ch 1 2 3 4
R5

H H H H

O = Press Option

S = Press Select