beautypg.com

頁面 52 – Elenco 30MHz Dual Trace User Manual

Page 52

background image

49

APPLICATION

MEASUREMENT OF VOLTAGE BETWEEN TWO POINT
ON A WAVEFORM

(Refer to Fig.8)

This procedure may be used to measure peak-to peak voltages, or for measuring the
voltage difference between any two points on a waveform.

1. Connect the signal to be measured to the input connector, set the CH 1/CH 2

switch to the channel to be used, and set the AC-GND-DC switch to AC. Set the
VOLTS/DIV and sweep TIME/DIV controls to obtain a normal display of the
waveform to be measured. The VARIABLE control must be set to CAL.

2. Using the X POSITION control, adjust the waveform position such that one of the

row points falls on a major horizontal graduation line.

3. Using the X POSITION control, adjust the second point to coincide with the center

vertical graduation line.

4. Measure the vertical distance between the two points( at least 3 divisions

desirable for best accuracy). Multiply the number of divisions by the setting of the
VOLTS/DIV control. If a probe is used, further multiply this by the probe
attenuation ratio.

The measurement is summarized by the following equation:

Voltage = Vert div x VOLTS/DIV x probe

100

90

10

0

Measuring point adjusted to the
center vertical scale by position

Ground potential
adjusted by POSITION
(reference line)

100

90

10

0

Adjust to the center
vertical scale with
POSITION

Points to be measured

Adjust to the horizontal
scale with POSITION

Fig.7 DC Voltage Measurement

Fig.8 Voltage Measurement

V

e

rt

ic

a

l d

is

ta

n

ce

V

e

rt

ic

a

l d

is

ta

n

ce

b

e

tw

e

e

n

tw

o

p

o

in

ts