Carrier 16JA User Manual
Page 9
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

TROUBLESHOOTING GUIDE (CONT)
SYMPTOM OR DIFFICULTY
POSSIBLE CAUSE
REMEDY
D. Machine shuts down on
safety control.
1. Motor overloads.
1. Reset all motor overloads and
check reason for failure.
2, Hermetic pump overload has
tripped.
2. a. If pump is out of liquid, add
solution or water.
b. If pump is solidified, de-
solidify. (Refer to Solution
Desolidification.)
3. Shutdown on low-temperature
cutout.
3. a. Check low-temperature cut
out setting.
b. Control point adjuster set
ting too low. Turn up chilled
water controller to design.
c. Check condensing water tem
perature control.
E. Solidification during
shutdown.
1, Dilution cycle not long enough.
1. Check setting of dilution ther
mostat, Should be 140F. If still
solidifies reset to lower cutout
temperature.
2. Improper closing of capacity
control valve.
2. Check valve closure, Desolidify,
3. Machine is shut down, but con
densing water pump is still
running.
3. Manually shut down condensing
water pump.
F, Suspect air leakage.
Leakage into vacuum side of
machine.
Determine noncondensable ac
cumulation rate. Refer to the
Carrier Standard Service Tech
niques, publication SM-16.
G. Loss of vacuum at
shutdown.
Leakage into vacuum side of
machine.
Leak test machine.
H, Failure to keep ma
chine purged.
1, Leakage above the pumping rate
of purge.
1. a. Perform leak rate check,
b. Leak test machine.
2. Purge not performing properly.
2. a. Valves not opened properly.
Check to see that all valves
are in correct position,
b. Purge solidified. Desolidify.
c. Lack of solution flow to purge
from solution pump. Con
tact
your
Carrier
repre
sentative.