General information
☞
Read and comply with this manual before installing or operating the equip-
ment.
☞
Do not use the equipment if it is damaged.
To operate the vacuum controller, valves and/or vacuum pumps are necessary. Oth-
erwise the controller works as measuring device.
Remove all packing material, remove the product from its packing-box, remove the
protective covers and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
Intended use
☞
Ensure that the individual components are only connected, combined and oper-
ated according to their design, or with OEM parts, and as indicated in the instruc-
tions for use.
☞
Comply with notes on correct vacuum and electrical connections, see section
☞
The equipment is designed for
ambient and gas temperatures at continuous
operation between +10°C and +40°C and up to 80°C for short periods at the
vacuum connection. Check the maximum temperatures, especially if installing
the equipment in a cabinet or a housing and make sure ventilation is adequate.
Ensure that the maximum permitted gas temperature at the pressure transducer
(see „Technical data“, pg. 12) is not exceeded.
Use the equipment and all system parts
for the intended use only, i.e. for control
and measurement of vacuum in vessels designed for that purpose.
Connecting the controller
☞
Do not permit any uncontrolled pressurizing (e.g., make sure that pipelines
cannot become blocked) to avoid a risk of bursting!
☞
Check the maximum ambient and gas temperatures (see „Technical data“, pg.
12, especially if installing the equipment in a cabinet or a housing or in case of
elevated ambient temperatures.
•
Comply with max. permitted pressure at the pressure transducer, see section
• Connect hoses gas tight at the vacuum connection of the pressure transducer.
• Ensure the stability of hose connections.
NOTICE
NOTICE
Safety information