Write commands compatible to ”CVC 3000”
** If remote operation is selected or deselected, the user has to ensure that no dangerous status of
the system due to the change of the mode of operation can occur and to provide appropriate safety
measures, especially if selecting remote operation interferes with a locally operated active process.
* Pressure setting with venting is only possible in operation mode ”Vacuum control”, if a venting valve is
connected and configurated and ”Vacuum control” is started. The venting valve opens automatically,
if the actual vacuum is 10 mbar below the preset vacuum. Automatic venting becomes inactive, if
”Vacuum control” is stopped (”Esc”), if setting a pressure value using the command OUT_SP_1 or if
the operation mode is changed. Activate the command OUT_SP_V again if necessary.
X
X
X
X
Function
operation mode
set vacuum
set vacuum
with venting*
speed
start-up pressure
delay
switch off pressure
switch off time
delay
Attention: If control
is running only
switching from 1 to
2 and 2 to 1 is pos-
sible.
Description
0: VACUULAN
1: Pump down
2: Vac control
4: RC 5/6 management
Parameter
X
XXXX
XXXX
XXX
XXXX
XX:XX
XXXX
XX:XX
Command
OUT_MODE
OUT_SP_1
OUT_SP_V
OUT_SP_2
OUT_SP_3
OUT_SP_4
OUT_SP_5
OUT_SP_6
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
speed in %, (100% = ”HI”)
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
hh:mm (hours:minutes)
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
hh:mm (hours:minutes)
00:00 = off
0 remote off
1 remote on
2 CVC 2000 commands
3 CVC 3000 commands
0 venting valve closed
1 venting valve open
store settings permanently
1 stop
2 stop with adoption of the set vacuum
START
STOP
REMOTE**
CVC
OUT_VENT
STORE