VACUUBRAND VNC 2 E User Manual
Page 31
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999199 / 28/11/2012
Read commands compatible to ”CVC 3000”
Function
status of
process control
fault status
0: VACUULAN
1: Continuous pumping
2: Vacuum control
4: RC 5/6 management
0: venting valve closed
1: venting valve open
0: coolant valve closed
1: coolant valve open
0: in-line valve closed
1: in-line valve open
0: pump off
1: pump on
Description
Response
XXXXXX
XXXXXXXXX
Command
IN_STAT
IN_ERR
0: last interface command correct
1: last interface command incorrect
0: no internal fault
1: internal fault (overtemperature at relais)
0: no external fault, catchpot empty
1: external fault, catchpot full
0: no fault at pressure transducer
1: fault at pressure transducer
0: no overpressure
1: overpressure
0: no fault at venting valve
1: fault at venting valve
0: no fault at coolant valve
1: fault at coolant valve
0: no fault at in-line valve
1: fault at in-line valve
0: no fault at pump
1: fault at pump
0: control off
1: pump down
2: set vacuum reached
3: vacuum below set vacuum