Application examples, Vacuum for filtration and suction, Assembly of a vacuum system – VACUUBRAND PC 620 NT plastic housing User Manual
Page 64
page 64 of 113
Assembly of a vacuum system
+
Assemble vacuum connection lines between controller, vacuum pump
(diaphragm pump with in-line valve or Vacuum-Management-System)
and vacuum application.
+
Assemble electrical connections.
+
Connect coolant if necessary.
Vacuum for filtration and suction
+
Select ”
Pump down” function.
+
Set ”Minimum” to a value which provides adequate suction but does
not lead to evaporation of the solvent.
+
Start process by pressing ”START/STOP” key.
alternatively:
+
Select ”
Vac control” function.
+
Set ”Set vacuum” (and ”Hysteresis”, if necessary) to a value which
does not lead to evaporation of the solvent.
+
Set ”Maximum” pressure setting so that pump will switch off at the end
of the filtration process, or in the event that a filter cake cracks, leading
to a sudden pressure increase.
+
Start process by pressing ”START/STOP” key.
+
Tip for filtration: Adjust preset pressure to a value well above the
boiling pressure of the solvent (e.g., for water >>15 Torr (20 mbar)).
Set the maximum pressure to e.g., 375 Torr (500 mbar). Once the fil-
tration has finished, the pressure increases and the pump is switched
off automatically.
Application examples