VACUUBRAND PC 620 NT plastic housing User Manual
Page 57
page 57 of 113
+
Delay: ”Delay” determines the time the pump (with VMS module and
in-line valve) remains running and the coolant valve remains open af-
ter the process has been stopped.
The ”Delay” is adjustable in a range of 1-300 minutes or can be set to
”Off” (”Off” means that when the process stops, the pump is switched
off and the coolant valve closes immediately.).
+
Duration: ”Duration” determines the total process time since control
start.
A preset ”Duration” (process time) has no effect if the process is
stopped due to a preset ”Maximum” before ”Duration” is reached.
The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be
set to ”Off”. ”Off” means that no endpoint of the process is defined.
The screen-shot shows the factory-set values.
Vac control
When selecting ”Graphic” the display shows a pressure vs. time curve.
The timeline in the diagram adapts automatically to the process time.
+
Press the selection knob twice to return to the standard display.
Vac control
Set vacuum 75 Torr
Hysteresis Auto
Maximum
Off
Delay
Off
Duration Off
- - - - - - Graphic - - - - - -
- - - - - - - Back - - - - - - -
760 Torr