cess will start immediately after a power failure with-
out pressing any further key. It is your responsibility to
ensure that automatic start-up of the system will not
lead to any dangerous condition. Provide appropriate
safety measures. Check
prior to starting the process
whether the option ”Autostart” (menu: configuration) is
enabled.
• Pay attention to the safety symbol ”hot surfaces” on
the equipment. Hot parts may cause burns if touched.
Adopt suitable measures to prevent any danger aris-
ing from hot surfaces or electric sparks. Ensure that
hot surfaces of the pump do not cause burns. Provide
a suitable contact guard if necessary.
• Ensure that the
coolant outlet pipe at the waste vapor
condenser is always free and that it cannot get blocked.
Prevent the backpressure of gases and the backflow of
condensates at the outlet.
Check the liquid level in the catchpots regularly and drain
condensate in time to prevent overfilling. Install a level
sensor (see ”Accessories”, pg. 86) for monitoring, if
necessary.
Provide appropriate protective measures to allow for the
possibility of failure and
malfunction. The protective mea-
sures must also allow for the requirements of the respec-
tive application.
The CVC 3000 controller is powered by a short circuit
proof wide-range power supply with integrated overload
protection.
In case of overload, the motor is shut down by a
self-hold
thermal cutout in the winding.
Note: Only manual reset is possible. Switch off the pump
and disconnect from the power source. Identify and elimi-
nate the cause of failure. Wait approx. five minutes before
restarting the pump.
NOTICE