Troubleshooting

Fault

Possible cause

Remedy

❑ No display or display

disappears.

➨ Automatic sleep mode?

✔ Switch on, increase opera-

ting time if necessary.

➨ Battery flat?

✔ Replace battery.

➨ Other cause?

✔ Contact local distributor.

❑ Incorrect pressure read-

ing.

➨ Pressure transducer not cor-

rectly adjusted?

✔ Readjust pressure trans-

ducer.

➨ Moisture in the pressure

transducer?

✔ Dry or evacuate the pressure

transducer and readjust if

necessary.

➨ Battery flat, battery symbol is

displayed?

✔ Replace battery.

❑ Display is flickering and/

or battery symbol ap-

pears.

➨ Battery flat?

✔ Replace battery.

❑ Adjustment mode cannot

be activated.

➨ At the applied pressure, read-

justment is not possible.

✔ Admit air to atmospheric

pressure or evacuate to a

pressure below 20 mbar.

❑ All segments of the LCD

are displayed or no dis-

play although battery has

been replaced.

➨ Pressure transducer or meas-

uring electronics defective?

✔ Contact local distributor.

❑ Pump does not start or

stops immediately.

➨ Mains cable not plugged in?

✔ Plug in mains cable. Check

mains fuse,

➨ Pressure in outlet pipeline too

high?

✔ Remove blockade in line,

open valve.

➨ Overload of motor?

✔ Allow motor to cool down,

identify and eliminate cause

of failure. Manual reset is

necessary. Switch off pump

or unplug mains.

Only C/US version:

Motor fuse blown?

✔ Check and replace motor

fuse. Identify and eliminate

cause of failure.