During operation, Notice – VACUUBRAND PC 301 User Manual
Page 21
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014
page 21 of 40
Separators
Assembling the hose nozzle with union nut at the inlet and at the outlet:
➨
Take the hose nozzle with attached compression ferrule and union nut out of the
round bottom flask and put onto inlet / outlet connection.
➨
Tighten the union nut by hand until you can feel the stop. Then tighten an addi-
tional 1/4 rotation with an open-ended wrench size 17 for final installation.
Catchpots:
The catchpot at the inlet prevents droplets and particles from
entering the pump.
☞
Lifetimes of diaphragms and valves are enhanced.
☞
Improves vacuum in case of condensation.
Use the condensate separator (catchpot AK) at the outlet.
☞
No draining of condensate back into the pump, controlled
recovery of condensate.
Both catchpots are coated with a protective layer to prevent
disintegration in case of breakage or implosion.
➨
Assemble the catchpots at the inlet and at the outlet using
joint clips.
During operation
➨
Vent and dispose of
potentially dangerous gases or vapours at the outlet of
the pump appropriately.
☞
Due to the high compression ratio, the pump might generate overpressure at the
outlet. Check pressure compatibility with system components (e.g., exhaust tub-
ing or exhaust valve) at the outlet. Ensure that the pump outlet is neither blocked
nor restricted.
☞
Maximum ambient temperature: 40 °C
Check the maximum temperatures, if installing the pump in a cabinet or a hous-
ing. Make sure ventilation is adequate, especially if the ambient temperature is
elevated.
• If the pump is installed at an altitude of more than 2000 m above mean sea
level, check compatibility with applicable safety requirements, and adopt suitable
measures. There is a risk of the motor overheating due to insufficient cooling.
• Check compatibility with the
maximally permitted pressure at outlet and the
maximum pressure difference between inlet and outlet ports.
Do not start the pump if the
pressure difference between inlet and outlet ports
exceeds max. 16.0 psi (1.1 bar). Attempts to start the pump at higher pressure dif-
ference may cause stalling and damage of the motor.
If pumping condensable vapours (water vapour, solvents, etc.), let the pump run with
gas ballast to help purge any condensation in the pump.
Prevent internal condensation, transfer of liquids or dust. The diaphragms and valves
will be damaged, if liquids are pumped in significant amounts.
Check the pump regularly for external soiling and deposits. Clean the pump if neces-
sary to avoid an increase of the pump’s operating temperature.
catchpot
at outlet
catchpot
at inlet
NOTICE