Replacing the battery, Accessories – VACUUBRAND PC 301 User Manual
Page 26
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014
page 26 of 40
Accessories
Vacuum hose (caoutchouc) 10 mm ID ......................................................................................686002
Silencer for hose nozzle DN 10 mm ............................................................................................636588
For additional accessories such as vacuum valves, small-flange components, vacuum gauges
or vacuum controllers refer to www.vacuubrand.com
Upgrade possibilities
Modification kit for small flange KF DN 16 at inlet ....................................................................699939
Elbow piece (90°) for PTFE tubing DN 10/8 mm for assembly at inlet .......................................637873
Solenoid operated valve (C3-B) for assembly at distribution head .............................................636668
Controller CVC 3000 ...................................................................................................................683160
In-line valve VV-B 6C ..................................................................................................................674291
Level sensor.................................................................................................................................699908
Elbow connecting piece (90°) for PTFE tubing DN 10/8 mm....................................................638434
T-piece for PTFE tubing DN 10/8 mm ........................................................................................638435
PTFE tubing DN 10/8 mm (sold by meter) ...................................................................................638644
Attention: Dust-laden gases, deposits and condensed solvent vapour can restrict air flow out the si-
lencer. The resultant back pressure can lead to damage of pump bearings, diaphragms, and valves.
Under those conditions, a silencer must not be used.
Note: Some of the components on the printed circuit board are sensitive to electrostatic discharge.
Do not touch components. If necessary, ground the printed circuit board and the person replacing the
battery suitably.
Attention: Use only a battery of the recommended type (see ”Technical data”).
The battery must not be charged or connected to an external voltage.
➨
Remove the battery.
➨
Insert the new battery in the battery compartment correctly
matching the positive and negative polarity (as shown by
the engraving inside the compartment).
☞
When reassembling the vacuum gauge, make sure that
the switch pad and the printed circuit board are positioned
correctly.
➨
Carefully screw both parts of the housing together.
battery
switch pad
printed circuit board
Replacing the battery
➨
Unscrew the four countersunk head screws at the rear
side of the housing by using a Phillips screw driver.
➨
Disassemble the housing parts carefully.