Technical data
Type
PC 301 / MD 1C + 2AK
Maximum pumping speed
at 50/60 Hz (ISO 21360)
m
3
/h
1.3 / 1.5
Ultimate vacuum (absolute) without gas ballast
mbar
2
Ultimate vacuum (absolute) with gas ballast
mbar
4
Maximum permissible inlet pressure (absolute)
bar
1.1
Maximum permissible outlet pressure (abso-
lute)
bar
1.1
Maximum pressure difference between inlet
and outlet
bar
1.1
Maximum permissible pressure (absolute) at
gas ballast valve
bar
1.2
Permissible ambient temperature
storage / operation
°C
-10 to +60 / +10 to +40
Permissible relative atmospheric moisture
during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
m
2000
Rated power
kW
0.08
No-load speed at 50/60 Hz
min
-1
1500 / 1800
Maximum permissible range of supply voltage
( ±10% )
Attention: Observe specifications of rating
plate!
200-230 V~ 50/60 Hz
120 V~ 60 Hz
Maximum rated current at:
120 V~ 60 Hz
200-230 V~ 50/60 Hz
A
A
1.7
0.8 / 0.85
Motor protection
thermal cutout, manual reset;
MD 1C C/US: additionally fuse 2.5 AT
Degree of protection IEC 529
IP 40
Inlet
hose nozzle DN 6/10 mm
Outlet
hose nozzle DN 6/10 mm
A-weighted emission sound pressure level*
(uncertainty K
pA
: 3dB(A))
db(A)
45
Dimensions L x W x H approx.
mm
316 x 239 x 460
Weight approx.
kg
10.2
We reserve the right for technical modification without prior notice!
*
Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and at ultimate vacuum with exhaust
tube at outlet.