When switching off the device, the remaining battery capacity

is displayed qualitatively by the number of arrows in the dis-

play for a few seconds.

If the battery symbol appears in the display during meas-

urement, the battery should be replaced.

The correct function of the device is no longer ensured.

Display of the status of discharge of the battery

Adjusting the operating time and the measuring cycle

Press key MODE to select the time setting mode.

The clock symbol and the actual operating time in minutes

are displayed.

To decrease operating time:

Press key MODE to change arrow direction to the left.

Press key UP/DOWN to decrease operating time.

Press key ON/OFF to confirm.

The device switches to mode ”

adjustment of measuring

cycle”.

The

measuring cycle can be chosen between Automatic,

1 measurement in 3s, 1 measurement in 1s and 3 meas-

urements in 1s.

The arrow to the right indicates that the time can be in-

creased by using key UP/DOWN.

Press key UP/DOWN to increase operating time.

The measuring cycle as from the last operation (factory-

set: ”A”) is displayed.

Press key UP/DOWN to select the measuring cycle: A

= Automatic (automatic determination of the measuring

cycle; frequent measurement in case of great pressure

change), S1 = 1 measurement in 3s, S2 = 1 measurement

in 1s, S3 = 3 measurements in 1s.

The flashing of the arrow (only ”S1”, ”S2” or ”S3”) symbol-

izes the frequency of the pressure measurement).

Press key MODE to change the direction of the adjust-

ment.

To adjust measuring cycle:

The vacuum gauge is equipped with an automatic sleep

mode to save battery power when the instrument is not

used. When the clock symbol appears on the display, the

vacuum gauge will switch off automatically after approx. 1

minute.

Press key ON/OFF, the clock symbol disappears and the

preselected operation time will start again.

Clock symbol

The time until

automatic switching off (operating time) can be set by the user within a range of 1

to 999 minutes (factory-set: 15 minutes) or

continuous operation ”On”.

Press key ON/OFF to confirm the measuring cycle and to

terminate mode.