Fittings and tubing of the different pump models – VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual
Page 64
page 64 of 76
Outlet side
MV 10C VARIO-B
PC 3010 VARIO
MD 12C VARIO-B
PC 3012 VARIO
Inlet side
Pumps VARIO-B /
Pumping units PC 301x VARIO
Fittings and tubing of the different pump models:
➨
Disassemble the housing covers (A) to
check the diaphragms and valves.
➨
Unscrew four Allen screws (R) with a 5mm
wide Allen key. Remove the housing cover
with housing cover insert (A) together with
the head cover (D), the valves (B) and the
O-rings (C) (head cover with valves and O-
rings will remain within the housing cover).
+
Never remove parts by using a pointed or sharp-edged tool (e.g., screw
driver). We recommend to use a rubber mallet or compressed air (to
be blown carefully into port).
➨
Remove the head cover (D) carefully from the housing cover. Note the
position and orientation of the valves (B) and remove them.
➨
Replace valves and/or O-rings if necessary.
+
Use petroleum ether or industrial solvent to remove deposits. Do not
inhale vapors.
R
- MV 10C VARIO-B without controller (o.C.) ME 4C NT VARIO without controller (o.C.) MD 4C NT VARIO without controller (o.C.) MD 12C VARIO-B without controller (o.C.) PC 3012 VARIO without controller (o.C.) PC 3010 VARIO without controller (o.C.) PC 3004 VARIO without controller (o.C.) PC 3003 VARIO without controller (o.C.) PC 3002 VARIO without controller (o.C.)