Replacing the valves – VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual
Page 57
page 57 of 76
Loosen the clamping brackets (S) on the valve
heads (O).
➨
Unscrew at each clamping bracket the two
countersunk screws with a Torx driver T20.
Remove the clamping brackets.
Loosen connection fastener slightly.
➨
Turn the fillister head screw (L) with a Torx
driver T20 at most one turn.
+
Do not detach the fillister head screw from
the square nut (K).
➨
Open the hinged cover of the connection
fastener (Q) with a slotted screwdriver.
Replacing the valves
Q
O
S
➨
Remove valve heads (O) along with the disc
springs (R), connection tube if applicable,
hose nozzles (P) and connection fasten-
ers (Q) or move the valve heads carefully
aside. Note position and orientation of the
valve heads.
+
Note position and alignment of valves (N).
➨
Check valves (N) and O-rings (M) for damage and soiling.
➨
Replace valves or O-rings if necessary.
➨
Use petroleum ether or other industrial solvent to remove deposits. Do
not inhale vapors.
O
R
N
Q
- MV 10C VARIO-B without controller (o.C.) ME 4C NT VARIO without controller (o.C.) MD 4C NT VARIO without controller (o.C.) MD 12C VARIO-B without controller (o.C.) PC 3012 VARIO without controller (o.C.) PC 3010 VARIO without controller (o.C.) PC 3004 VARIO without controller (o.C.) PC 3003 VARIO without controller (o.C.) PC 3002 VARIO without controller (o.C.)