Troubleshooting – VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual
Page 48

page 48 of 76
Troubleshooting
Fault
Possible cause
Remedy
❑ Pump does not
start or stops im-
mediately.
➨ Pump has been ex-
posed to condensate?
✔ Allow pump to run for
some minutes at maxi-
mum speed with atmo-
spheric pressure at the
inlet.
➨ Overpressure in outlet
line?
✔ Remove blockage in line,
open valve.
❑ Pump does not
achieve its ulti-
mate vacuum or
usual pumping
speed.
➨ Centring ring at small
flange connection not
correctly positioned, or
leak in the pipeline or
vacuum system?
✔ Check pump directly -
connect vacuum gauge
directly at pump inlet -
then check connection,
pipeline and vacuum
system if necessary.
➨ Long, narrow vacuum
line?
✔ Use lines with larger di-
ameter, length as short as
possible.
➨ Pump has been exposed
to condensate?
✔ Allow pump to run for
some minutes with atmo-
spheric pressure at the
inlet to purge.
➨ Deposits have been
formed inside the pump?
✔ Clean and inspect the
pump heads.
➨ Diaphragms or valves
damaged?
✔ Replace diaphragms and/
or valves.
➨ Outgassing substances
or vapor generated in
the process?
✔ Check process parame-
ters.
➨ Pump too hot?
✔ Allow pump to cool down.
Determine and eliminate
the cause of overheating.
❑ Pump too noisy.
➨ Atmospheric or high
pressure at the pump
inlet?
✔ Connect hose or silencer
to pump outlet. Be careful
not to cause outlet over-
pressure, especially with
condensable vapors.
➨ Diaphragm crack or
diaphragm clamping disc
loose?
✔ Perform maintenance.
➨ Other than above men-
tioned causes?
✔ Contact local distributor.
❑ Pump seized.
✔ Contact local distributor.
- MV 10C VARIO-B without controller (o.C.) ME 4C NT VARIO without controller (o.C.) MD 4C NT VARIO without controller (o.C.) MD 12C VARIO-B without controller (o.C.) PC 3012 VARIO without controller (o.C.) PC 3010 VARIO without controller (o.C.) PC 3004 VARIO without controller (o.C.) PC 3003 VARIO without controller (o.C.) PC 3002 VARIO without controller (o.C.)