VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual
Page 60

page 60 of 76
➨
Put head cover cowling on.
➨
Slide the head cover cowling in the grooves
of the caps (I) and under the connection
fasteners (Q).
➨
Install the washers. Use a Torx driver T20 to
attach the 4 screws holding the head cover
cowling.
➨
Tighten the fillister head screws (L) of the
connection fasteners (Q) with a Torx driver
T20.
➨
Close the hinged covers.
Replace diaphragms and valves of the opposite side of
the pump in the same way.
➨
Feed the PTFE hose into the inlet of the
exhaust vapor condenser. Mount the con-
denser with the counter holder (Torx screws
T10) to the pump. Fasten union nut.
PC 3002/3003/3004 VARIO:
➨
Check the overpressure safety relief device
at the exhaust waste vapor condenser.
➨
If necessary, pull off the old pressure relief
valve and install the new one. Check for
correct position of the PTFE-foil under the
pressure relief valve.
- MV 10C VARIO-B without controller (o.C.) ME 4C NT VARIO without controller (o.C.) MD 4C NT VARIO without controller (o.C.) MD 12C VARIO-B without controller (o.C.) PC 3012 VARIO without controller (o.C.) PC 3010 VARIO without controller (o.C.) PC 3004 VARIO without controller (o.C.) PC 3003 VARIO without controller (o.C.) PC 3002 VARIO without controller (o.C.)